96-channel centering on row-layout labware (Flex 9.1.2)

Suggested protocol smoke for #21803 (96-channel partial-row centering over nest_8_reservoir_22ml v2). Requires a 96-channel pipette; not runnable on a Flex configured with only P50 single/multi.

Monorepo branch: chore_release-9.1.2 · compared to v9.1.1 · robot OS 9.1.2-alpha.0

Hardware: flex_96_channel, nest_8_reservoir_22ml

Related PRs

Harness commands

uv run flex-test inspect
uv run flex-test probe --no-picture

Tests

IDPlan
C1
Partial-row aspirate/dispense centering
blocked

96ch partial row configs must center on 8-row reservoir wells.

  • Attach Flex 96-channel pipette.
  • Load nest_8_reservoir_22ml (definition v2) on deck.
  • Run a short protocol with partial-row configuration aspirate/dispense.
  • Visually confirm nozzles center on target wells (no edge offset).

KansasFLEX currently has P50 multi + P50 single only

C2
Compare against pre-9.1.2 if available
skipped

Confirms geometry change vs prior robot OS.

  • Optional A/B on a second robot or prior build snapshot.